NIH3D

MRI-Visible Laterality Marker STL

Created by
Scott Robertson
Created:
7/10/25
Submitted:
10/7/26
Published:
10/7/26

Select an image below to view

Learn more about NIH 3D workflow outputs

3DPX-022785

Licensing:

CC-BY
213
77
Version 2

Category

Hardware & Devices
Devices and Hardware
Description

These files contain 3-D-printable Right and Left external laterality markers for use during MRI examinations.


Background

Conventional MRI relies on technologist setup and accuracy of DICOM metadata to indicate patient side. The lack of a visual confirmation can result in wrong-side labeling errors. We designed low-profile silicone markers, doped with 0.2 % gadobenate dimeglumine (MultiHance), that appear conspicuously bright on all routine localizer sequences without introducing artifacts or additional SAR penalty. The hard PLA shell makes these markers easily cleanable and reusable. Each marker costs less than a dollar to produce.


Contents

smallLateralityMold_L.stl - Mold shell for left laterality marker (approx. 40 mm x 40 mm x 10 mm)

smallLateralityMold_R.st - Mold shell for right laterality marker (approx. 40 mm x 40 mm x 10 mm)

smallLateralityTop_L.stl - Top for left laterality marker mold (approx. 40 mm x 40 mm x 10 mm)

smallLateralityTol_R.stl - Top for right laterality marker mold (approx. 40 mm x 40 mm x 10 mm)


Printing & assembly

  1. Print in PLA or PET-G.
  2. Prepare two-part RTV silicone; dope with 0.2 % v/v gadobenate dimeglumine (MultiHance®).
  3. Degas the silicone in a vacuum chamber
  4. Fill the mold with silicone, attach the tops, and cure 24 h.
  5. Wipe off excess silicone


License

Released under CC BY-4.0 – free for clinical, research, and commercial use with attribution.


Citation

Robertson SH et al. “Improving Patient Safety with MRI-Visible Laterality Markers.” AAPM 67th Annual Meeting, 2025.