NIH3D

Challenge Residue Removal Assembly (CRRA) - 10mm sample wedges

Created by
AMMP
Created:
5/14/18
Submitted:
3/6/23
Published:
3/6/23

Select an image below to view

3DPX-009147

Licensing:

GNU GPLv3
93
6
Version 2

Category

Hardware & Devices
Devices and Hardware
Description

These are example porous wedges created from the blank files in the blank 10 mm CRRA entry in the NIH 3D Print Exchange. For full description and use instructions visit https://3dprint.nih.gov/discover/3dpx-009146.

The Challenge Residue Removal Assembly (CRRA) is a research tool designed by OSEL FDA researchers to mimic medical devices, namely orthopaedic, to evaluate challenges in removing additive manufacturing residual materials from porous geometries.

Ten-millimeter tall wedges are provided with three size variations of the same unit cell. All porous structures were created with circular struts. Each file name indicates the wedge height and nominal pore size as follows: Nmm_Pum, where N is the height of the corresponding assembly disc and P is the nominal pore size. All porous structures were created with circular struts. Each file name indicates the wedge height and nominal pore size as follows: Nmm_Pum, where N is the height of the corresponding assembly disc and P is the nominal pore size.

Nominal pore size
[µm]

Strut diameter
(D) [µm]

Unit cell size (A=B=C) [mm]

Porosity
 (%) [-]

700

700

1.979

54%

1000

700

2.404

66%

1250

700

2.758

73%

 

The provided wedges do not represent a definitive challenge or test for any specific medical device.